7 programmable function keys, 8 configuration (set-up) – Bowers Group Trimos Mestra Height Gauges User Manual

Page 19

Advertising
background image

750 50 0001 03

19/28

7 Programmable function keys

The operating handle of the instrument includes one key to
activate the air cushion displacement of the instrument (Air)
and 2 programmable function keys, one yellow key (1) and
one blue key (2). The pre-programmed functions of this 2
keys are as follows :

1. Yellow key:

Preset

2. Blue key:

Functions


Other functions can be allocated to these keys, see § 8.

8 Configuration

(Set-up)

Certain parameters of the instrument can be configured
according to the user requirements.
To open this Set-up menu, press the Functions key longer
than 2 seconds.

> 2 sec.

The cursor keys allow the selection of the different elements
in the menu. The yellow arrow indicates the cursor position.

The Enter key allows to activate / cancel the configuration
fields.

For selection of the next page, press the Functions key,

To quit the Set-up menu, press the Enter key longer than 2
seconds.

> 2 sec.

Page 1

1. Locking of the measuring unit (mm or inch):

By activating one or the other field, the mm/inch key will
be locked in for one or the other measuring unit. The
instrument display will than only work in the selected unit.

2. Height of the setting gauge :

If another setting gauge than the supplied one will be
used, it is possible to enter its height.
The checking and storage of the constant must always be
done between 2 inverted surface without moving the
instrument or the reference gauge.


3. Angle unit format:

x.x°: Decimal

degrees

x° x' x": Degrees, minutes, seconds
rad: Radians

4. Height of the gauge block for measuring angles:

By activating the required field, the last 2 steps of the
angle measurement will be cancelled. The height of the
gauge block will be taken over.

1

2

1

2

3

4

Advertising