2 Introduction

930 Compact IC Flex Oven (2.930.2100)

■■■■■■■■

3

Inline filter
Inline filters protect the separation column reliably from potential contami-
nation from the eluent. The small filter pads with 2 µm pore size can be
replaced quickly and easily. They remove particles from the solutions, such
as bacteria and algae.

Pulsation absorber
The pulsation absorber protects the separation column from damage
caused by pressure fluctuations when switching the injection valve, and
reduces interfering pulsations during highly sensitive measurements.

Injection valve
The injection valve connects the eluent path to the sample path. By a
quick and precise switching of the valve a quantity of sample solution
defined by the size of the sample loop is injected and flushed to the sepa-
ration column with the eluent.

Detector
Metrohm offers a series of different detectors for various analysis tasks. A
suitable detector type must be ordered as a separate accessory.

Separation column
The intelligent separation column separates different components accord-
ing to their interactions with the column. Metrohm separation columns
are equipped with a chip where their technical specifications and history
(start-up, operating hours, etc) are stored.

2.2

Intended use

The 930 Compact IC Flex Oven is used for the determination of cations
and anions without suppression using ion chromatography.

The present instrument is suitable for processing chemicals and flammable
samples. Usage of the 930 Compact IC Flex therefore requires the user to
have basic knowledge and experience in handling toxic and caustic sub-
stances. Knowledge with respect to the application of the fire prevention
measures prescribed for laboratories is also mandatory.