Metrohm 807 Dosing Unit User Manual

Page 50

Advertising
background image

6.2 Dosing accuracy

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

807 Dosing Unit

Table 4

Permissible limit values as per ISO/EN/DIN 8655-3

Cylinder
volume
(mL)

max. systematic

Measurement deviation

max. permissible

Measurement deviation

2

± 0.5 %

± 10 µL

± 0.1 %

± 2 µL

5

± 0.3 %

± 15 µL

± 0.1 %

± 5 µL

10

± 0.2 %

± 20 µL

± 0.07 %

± 7 µL

20

± 0.2 %

± 40 µL

± 0.07 %

± 14 µL

50

± 0.2 %

± 100 µL

± 0.05 %

± 25 µL

The Metrohm agents worldwide offer the possibility of on-site dosing unit
inspections and certifications with respect to accuracy. We recommend
that an accuracy inspection be performed when the dosing cylinders and
dosing pistons of a dosing unit are replaced.

Advertising