Solutions, Viva «method, Method run – Metrohm 894 Professional CVS semiautomated User Manual

Page 5: Measuring commands

Advertising
background image

Application Bulletin 401/1 e

Installation instruction for 894 Professional CVS semiautomated

Dosing rate port 1

Maximum

mL/L

Dosing rate port 2

Maximum

mL/L

Dosing rate Fill port

Maximum

mL/L

Dosing rate special port

Maximum

mL/L

Tubing parameters

Port

Length

Diameter

Dosing port 1

Port 1

138 cm

2 mm

Dosing port 2

Port 3

0 cm

2 mm

Fill port

Port 2

25 cm

2 mm

Special port

Port 4

0 cm

2 mm

Valve disk

Rotating direction

automatic

3.3.2. Dosing unit at 800 Dosino (2), 800 Dosino (3) and

800 Dosino (4)

Dosing Unit name as used in the viva method templates.

800 Dosino (MSB 2)

894/D2 2mL Brightener

800 Dosino (MSB 3)

894/D3 2mL Suppressor

800 Dosino (MSB 4)

894/D4 2mL Std or sample

Parameters for preparation

Dosing port Prep/empty

Dosing port 2

Dosing rate port 1

2.0

mL/L

Dosing rate port 2

Maximum

mL/L

Dosing rate Fill port

Maximum

mL/L

Dosing rate special port

Maximum

mL/L

Tubing parameters

Port

Length

Diameter

Dosing port 1

Port 1

80 cm

0.3 mm

Dosing port 2

Port 3

0 cm

2 mm

Fill port

Port 2

55 cm

2 mm

Special port

Port 4

0 cm

2 mm

Valve disk

Rotating direction

automatic

3.4.

Solutions

Solutions that should automatically be dosed have to be

defined in the viva «Configuration» and need to be

assigned to the Dosing Unit which is used for the dosing.

The following table shows the solution names and assigned

Dosing Units as used in the viva method templates.

Solution name

Dosing Unit

VMS

894/D1 50mL VMS

Brightener concentrate

894/D2 2mL Brightener

Suppressor concentrate

894/D3 2mL Suppressor

Std or sample

894/D4 2mL Std or sample

S

OLUTION TYPE

for «VMS» and «Suppressor concentrate»

has to be A

UXILIARY SOLUTION

, for «Brightener concentrate»

and «Std or sample» it is S

TANDARD SOLUTION

.

For a step by step description please see the Tutorial

chapter «5.1.4 Define solutions».

4. viva «Method»

viva software includes three method templates for CVS

analysis with a semiautomated system.

Brightener determination (CVS, MLAT), semiautomated

Conditioning (Cu VMS), semiautomated

Suppressor determination (CVS, DT), semiautomated

The templates already establish the basic sequences for the

different applications and the use of 800 Dosinos for

automatic dosing as well as the calculation of the result.

However certain commands and settings have to be

adapted to the used hardware and the requirements of the

specific application.

For a step by step description please see the Tutorial

chapter «5.2 Methods for the semiautomated

determination».

The following commands have to be adapted:

4.1.

Method run

4.1.1. Measuring commands

Such as

Assign the 894 Professional CVS instrument to be used

on the tab «General/Hardware».

Adapt the measuring parameters on the tabs «Pre-

treatment», «Sweep», «Post-treatment» and

«Potentiostat» according to the requirements of the

application. These parameters can be found in separate

application documentation.

Page 5 of 7

Advertising