3.2 Installing the rinsing solution flow path

■■■■■■■■■■■■■■■■■■■■■■

8

■■■■■■■■

IC equipment: Liquid Handling Station

8 Inserting the adsorber tube

Insert the adapter into the SGJ opening of the bottle cap.

Screw the adsorber tube into the upper opening of the adapter.

Installing the capillary between the supply bottle and the
trap column

Required accessories

PEEK capillary (6.1831.180)

Metrosep I Trap 1 - 100/4.0 (6.1014.200)

Pressure screw (6.2744.010)

1 Removing the stoppers

Remove the stoppers from the trap column's inlet and outlet.
Keep the stoppers.

2 Installing the capillary to the column inlet

Tighten the other end of the installed capillary to the trap column
inlet using a pressure screw.