4.18 Conditioning
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
930 Compact IC Flex Oven/Deg (2.930.2160)
■
Visually inspect whether all capillaries and their connections from
the high-pressure pump to the detector are leak-tight. If eluent is
leaking out anywhere, tighten the corresponding pressure screw
further, or loosen the pressure screw, check the end of the capil-
lary and shorten it using the capillary cutter if necessary and
retighten the pressure screw.
4 Conditioning the system
Rinse the system with eluent until the required stability of the base-
line is attained (normally 30 minutes).
The instrument is now ready for measuring samples.