2 intended use, Intended use – Metrohm 940 Professional IC Vario ONE/SeS/PP/HPG User Manual

Page 13

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

940 Professional IC Vario ONE/SeS/PP/HPG (2.940.1540)

■■■■■■■■

3

Peristaltic pump
The peristaltic pump is used for pumping sample and auxiliary solutions. It
can rotate in both directions.

Metrohm CO

2

Suppressor (MCS)

The Metrohm CO

2

Suppressor (MCS) removes the CO

2

from the eluent

flow. This lowers the background conductivity, improves detection sensi-
tivity and minimizes the injection peak and carbonate peak.

Detector
Metrohm offers a series of different detectors for various analysis tasks. A
suitable detector type must be ordered as a separate accessory.

Sample degasser
The sample degasser removes gas bubbles and dissolved gases from the
sample.

High-pressure gradient module (HPG)
The second high-pressure pump in the instrument allows a gradient to be
created from two eluents. The software is used to control the amount of
eluent.

Separation column
The intelligent separation column separates different components accord-
ing to their interactions with the column. Metrohm separation columns
are equipped with a chip where their technical specifications and history
(start-up, operating hours, etc) are stored.

1.2

Intended use

The 940 Professional IC Vario ONE/SeS/PP/HPG is used for the determina-
tion of anions or polar substances using sequential suppression using ion
chromatography when the complex separation problem requires the use
of gradients.

Sequential suppression consists of:

Chemical suppression with a Metrohm Suppressor Module (MSM) and
subsequent

CO2 suppression with the Metrohm CO2 Suppressor (MCS).

Background conductivity is reduced to a minimum with sequential sup-
pression.

The second high-pressure pump in the lower plug-in allows for the con-
trolled mixing of two eluents.