9.1 Measuring interface

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

904 Titrando

9 Technical specifications

9.1

Measuring interface

Depending on the version, the 904 Titrando has one (version 2.904.0010)
or two (version 2.904.0020) galvanically isolated measuring interface(s).

The measuring cycle is 100 ms for all measuring modes.

9.1.1

Potentiometry

One high-ohm measuring input (Ind.) for pH, metal or ion-selective elec-
trodes and one measuring input (Ref.) for separate reference electrodes.

Input resistance

> 1 · 10

12

Ω

Offset current

< 1 · 10

–12

A (under reference conditions)

Measuring mode
pH

Measuring
range

–13 - +20 pH

Resolution

0.001 pH

Measuring
accuracy

±0.003 pH
(±1 digit, without sensor error, under reference conditions)

Measuring mode
U

Measuring
range

–1,200 - +1,200 mV

Resolution

0.1 mV

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)

9.1.2

Temperature

A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.

R (25 °C) and B value can be configured for NTC sensors.

Measuring range

Pt1000

–150 - +250 °C

NTC

–5 - +250 °C
(R (25 °C) = 30,000

Ω and B (25/50) = 4,100 K)