4.12 Suppressor

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

940 Professional IC Vario ONE/SeS/LPG (2.940.1450)

4.12.2

Connecting the suppressor

The three inputs and outputs of the suppressor units, numbered with 1, 2
and 3 on the connecting piece, each have 2 fixed mounted PTFE capilla-
ries.

Figure 4

Suppressor – connection capillaries

1

out
Output capillary for the eluent.

2

in
Input capillary for the eluent.

3

regenerant
Input capillary for the regeneration solution.

4

waste reg.
Output capillary for the regeneration solu-
tion; to the waste container.

5

waste rins.
Output capillary for the rinsing solution; to
the waste container.

6

rinsing solution
Input capillary for the rinsing solution.