4.2 Operating modes

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

Power Box

The Power Box is charged no matter whether the 899 Coulometer is
switched on or switched off.

899 Coulometer with the Power Box without power failure
mode

The power supply unit is connected to the Power Box and to the mains
supply.

The Power Box is switched off.

The 899 Coulometer is connected to the Power Box.

The 899 Coulometer is switched on.

The 899 Coulometer can be operated normally as long as the mains sup-
ply is stable. The Power Box does not take over supplying power if the
mains supply fails. The 899 Coulometer is shut down in a controlled way.

899 Coulometer with the Power Box without power supply
unit

The Power Box is fully charged and switched on.

The Power Box is not connected to the power supply unit and the
mains supply.

The 899 Coulometer is connected to the Power Box.

The 899 Coulometer is switched on.

The 899 Coulometer can be operated for at least 12 h (see Chapter 4.4,
page 12)
.

Stand-alone operation

The power supply unit is connected to the Power Box and to the mains
supply.

The Power Box is switched on.

The Power Box is charged. Once it is fully charged, the LED switches to a
higher light intensity (1-1).

The intelligence of the Power Box prevents overcharging, even if the
Power Box is connected to the mains supply for more than 14 h.