4 Functioning

800 Dosino

■■■■■■■■

19

The dosing drive sitting on the dosing unit turns, with the rotating drive
pin, the centering tube and thus the entire interior cylinder unit, with dos-
ing cylinder, cylinder base and the valve disc mounted within.

The bore hole of the valve disc faces a different distributor disc opening
after one rotation of the cylinder unit. This means that a different port is
selected for dosing (or filling).

4.3

Port assignments

The distributor of a dosing unit has four inputs/outputs (ports), which can
be freely selected. An additional connection, the VENT -Port (see Figure
12, page 19)
, leads directly to the underside of the distributor and can-
not be actuated by the valve disc. It deaerates the supply bottle and can
be equipped with an adsorber tube.

Port 1

Port 2

Port 3

Port 4

Vent

Figure 12

807 Dosing Unit from below

All ports of the 807 Dosing Unit can be used differently (see Chapter 4.4,
page 20)
. This is an important precondition for complex liquid handling
tasks. Titrandos use a standard port allocation which is the most suitable
for titration tasks.