Metrohm TitrIC pro III User Manual

Page 12

Advertising
background image

Application Bulletin 355

Installation Instructions for TitrIC pro III

Page 12 of 14

In tiamo

TM

it is typically used as a single determination

in autorepeat mode. In other words, it is always ready
to do a titration as soon MagIC requires it. To set it into
autorepeat go to the Run window, right-mouse-click and
select

Properties Run window

. Under Process select

Autorepeat determination

. Close the window with

<ok>

.

TitrIC pro III_pH calibration manual:

Here the iAquatrode

is taken out of the titration vessel and placed into the
buffer solutions. Start this method only from tiamo

TM

!

TitrIC pro III_titer determination of HCl:

This method is

used for the titer determination. It has to be used in
combination with the MagIC Net method

TitrIC pro

III_titer determination of HCl

.

TitrIC pro III_conductivity calibration:

This method is

used for calibration of the conductivity sensor. It has to
be used in combination with the MagIC Net method

TitrIC pro III_calibration conductivity

.

Import: Click on

Method

at the left (in MagIC Net and

tiamo

TM

), then select under

File / Method manager… / Edit /

Import…

Import the methods related to TitrIC pro III with the

Dilution kit from the MagIC Net CD. You will find the
methods in the folder example/TitrIC of the CD

.

Note!

Alterations of these methods should only be carried out by a
person at the administrator level and who is thoroughly
familiar with tiamo

TM

and MagIC Net

With the TitrIC pro – Combo Kit Dilution following analysis
can be carried out using the TitrIC pro III_dilution methods
in MagIC respectively tiamo

TM

:

The whole procedure is controlled by the MagIC Net
software which is the master software. In tiamo

TM

the

method

TitrIC pro III

is selected as a single determination.

Start the determination in autorepeat mode. To set it into
autorepeat go to the Run window, right-mouse-click and
select

Properties Run window

. Under Process select

Autorepeat determination

. Close the window with

<ok>

. Start

the determination. Now tiamo

TM

is waiting for the start

signal from MagIC Net via RS232.

In the MagIC Net software select the method

TitrIC pro

III_dilution

. Enter the required sample position, sample

identification etc. into MagIC Net. For value 1 choose
according to the list below:

Value 1 = 1 complete analysis including titration and
ion chromatography.

Value 1 = 2 only an ion chromatography analysis (e.g.
a calibration)

Value 1 = 3 only a titration analysis

With Value 1 = 1 a complete analysis is carried out.
Depending on the entered dilution factor the following will be
done:

Dilution Factor = 1:

Depending on the measured conductivity an
appropriate dilution factor is automatically applied by
MagIC Net. The calculation factor is calculated
according to Time Program No. 2. The finally used
dilution factor is the integer of the conductivity divided
by 250.

Dilution Factor >1:

If a dilution factor is known it can be directly entered into the
sample table. It will be diluted accordingly. The conductivity
measurement has no influence on the dilution in this case.

E.g. With Value 1 = 2 an ion chromatographic analysis will
be carried out with the entered dilution factor. No dilution
factor will be calculated as no conductivity measurement will
take place.