3 connecting the 801 stirrer, Connecting the 801 stirrer – Metrohm 801 Magnetic Stirrer User Manual
Page 17
Advertising
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
801 Stirrer
■■■■■■■■
9
Attaching the clamping ring and the electrode holder
6.2021.020
6.2013.010
Fasten the clamping ring and the electrode holder as follows:
1 Push the 6.2013.010 clamping ring onto the support rod.
2 Attach the 6.2021.020 electrode holder on the support rod.
3 Keep the red button pressed down and push the electrode holder
downwards to the required height.
4 Fix the position of the electrode holder with the aid of the clamping
ring.
3.3
Connecting the 801 Stirrer
Connect the magnetic stirrer as follows:
1 Exit the control software.
2 Connect the connection cable of the magnetic stirrer to one of the
sockets marked with MSB on the rear of the control device.