3 Learning by doing

IC Metrodata for Win95

48

Select the IC system type in the

Initial

tab. With

Anion

and

Cation

temperature coefficient and polarity are set automatically, with

User

these values can be combined independently.

Change the parameters for 732 IC Detector, 733 IC Separation

Center and 709 IC Pump to the values to be set for these instru-

ments at the start of the method.

First open the

Program

tab, then open the

Text

tab. Click the

Add

button to add a new program line.

Enter the time

x.x

in min into the first field of the program line.

Select the desired program point in the second field.

Select the parameter belonging to the program point in the third

field. For

Flow

,

Pmax

and

Pmin

values must be entered, for

Remote

1

,

0

or

*

must be entered. A certain remote line position can be

changed by moving the cursor in front of this position and then en-

tering

1

,

0

or

*

.

Click the

Add

button to add new program lines or click the

Delete

button to delete program lines. Faulty program points are dis-

played in the lower part of the program field.

If you wish to use predefined remote instructions, open the

Remote

configuration

tab.

Click the

Add

button to add a new remote instruction.

Enter any name you like for the remote instruction (max. 19 charac-

ters, e.g.

Ext. Valve Fill

) in the first field.

Enter

1

,

0

or

*

for the corresponding remote instruction for remote

control lines 1...8 in the second field. A certain remote line position

can be changed by moving the cursor in front of this position and

then entering

1

,

0

or

*

.

Click the

Add

button to add more remote instructions or click the

Delete

button to delete remote instructions.

Click the

Activate

button. The entered remote instruction are added

to the standard program points and can also be selected.

Open the

Text

tab and add the predefined remote instructions into

the time program.

After finishing the programming, click the

Apply

button. The pro-

gram and the parameters are tested for consistency.

Switch on the

ENABLED

option and close the window with

OK

.

Close the

Equipment

window with

Close

.

Select

Save method as

of the

File

menu and save the changed

method under the same or a new name.