Metrohm NH3-selective electrodes User Manual

Page 18

Advertising
background image

6.5 Performing measurements with very low concentrations

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

NH

3

-selective electrodes

Buret volume

10 mL

Sample size

10 mL

ISA/TISAB

10 mL

Total volume

20 mL

Factor c

Standard solution

/c

Sample

20

This results in a sample concentration in the measuring solution of 2.5
mg/L. The optimum concentration of the standard solution is thus 2.5
mg/L · 20 = 50 mg/L. Please note that this is merely to be considered a
guideline for standard additions. Even if you deviate from this recommen-
dation, precise measurements will still be possible.

6.5

Performing measurements with very low concentra-
tions

NOTE

The electrode 6.0506.100 or the membrane module 6.1255.000 must
be used for measurements with very low concentrations.

The response time of the electrode is relatively long with low ammonia
concentrations. The response time decreases with increasing concentra-
tion. You can improve the response time of the electrode and the preci-
sion of the result for ammonia concentrations < 6 · 10

–5

mol/L with the

following measures:

Condition the NH

3

-selective electrode prior to the measurement in an

ammonia-free pH 4 buffer solution.

Dilute the measuring electrolyte 1 : 9 with distilled water.

Work with sealed measuring vessel.

Use as large a sample volume as possible so that the ratio of surface to
volume is as low as possible. This will minimize the absorption of
ammonia from the air.

Even when these measures are applied, the response time of the electrode
may still be as long as ten minutes, not only for the samples but also
when conditioning in pH 4 buffer.

Advertising