14 injection valve, 1 connecting the injection valve, Connecting the injection valve – Metrohm 850 Professional IC Anion MCS Prep 1 User Manual

Page 53: Figure 24, Injection valve – connected, See chapter 2.14

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 Installation

850 Professional IC – Anion – MCS – Prep 1

■■■■■■■■

43

4 Guide the other end of the connection capillary out of the instrument

through a capillary feed-through.

Caution

If the sample degasser is not used, the input and output must be
sealed with the 6.2744.220 thread stoppers.

2.14

Injection valve

The injection valve connects the eluent and sample path. Through rapid
and precise valve switchover a precise amount of sample solution defined
by the size of the sample loop is injected and rinsed with eluent onto the
separation column.

2.14.1

Connecting the injection valve

The injection valve has six connectors: two for the sample path (connec-
tors 1 and 2), two for the eluent path (connectors 4 and 5) and two for
the sample loop (connectors 3 and 6).

Note

The capillaries of the eluent path and the sample path and the sample
loop are already installed in the newly delivered instrument.

1

2

3

4

5

6

1

2

3

4

5

6

7

7

7

7

Figure 24

Injection valve – connected

1

Injection valve

2

Sample loop
Connected to connectors 3 and 6.