4 installation, 1 setting up the instrument, 1 packaging – Metrohm 930 Compact IC Flex Deg User Manual

Page 18: 2 checks, 3 location, 2 removing transport locking screws, Setting up the instrument, Packaging, Checks, Location

Advertising
background image

4.1 Setting up the instrument

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

930 Compact IC Flex Deg (2.930.1160)

4 Installation

4.1

Setting up the instrument

4.1.1

Packaging

The instrument is supplied in highly protective special packaging together
with the separately packed accessories. Keep this packaging, as only this
ensures safe transportation of the instrument.

4.1.2

Checks

Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.

4.1.3

Location

The instrument has been developed for operation indoors and may not be
used in explosive environments.

Place the instrument in a location of the laboratory which is suitable for
operation, free of vibrations, protected from corrosive atmosphere, and
contamination by chemicals.

The instrument should be protected against excessive temperature fluctu-
ations and direct sunlight.

4.2

Removing transport locking screws

To avoid damage to the drives for the high-pressure pump and the vac-
uum pump during transport, the pumps are secured with transport lock-
ing screws. These are located at the rear of the instrument and labeled
with Transport security screws.

Remove these transport locking screws before the initial start-up.

Accessories

For this step you need:

Hex key 4 mm (6.2621.030)