6 Appendix

867 pH Module

■■■■■■■■

31

Inputs

+5 V

approx. 50 k

Ω Pull-up

tp

t

p

> 20 ms

active = low, inactive = high

Outputs

Open Collector

tp

t

p

> 200 ms

active = low, inactive = high

I

C

= 20 mA, V

CEO

= 40 V

+5 V: maximum load = 20 mA

The following tables offer information concerning the assignment of the
individual pins and their function:

Table 1

Inputs and outputs of the remote interface

Assignment

Pin No.

Function

*

Input 0

21

Start

Input 1

9

Stop

Input 2

22

Input 3

10

Quit

Input 4

23

Input 5

11

Input 6

24

Input 7

12

Output 0

5

Ready

Output 1

18

Conditioning OK

Output 2

4

Determination

Output 3

17

EOD

Output 4

3

Output 5

16

Error

Output 6

1

Output 7

2

Warning