Q45LMA Analog Output Card

Banner Engineering Corp.

Minneapolis, U.S.A.

www.bannerengineering.com • Tel: 763.544.3164

Explanation of Programmable Output Functions:

Switch 1:

Output Slope Select

On = Direct = Output value (voltage or current) increases with increasing distance

of the target from the sensor

Off* = Inverse = Output value decreases with increasing distance of the target

from the sensor

Switch 2:

Output Mode Select

On = The 4 to 20mA current output (white wire) is enabled
Off* = The 0 to 10V dc voltage output (black wire) is enabled

This switch configures the D/A driver to use either the current output or the voltage
output driver. This output function can only be set with the power to
the sensor turned off.

Switch 3:

Loss of Echo Mode Select

On = Min - Max Mode
Off* = Hold Mode

This switch determines the output response to the loss of echo. The “Hold Mode”
(Switch 3 Off*) maintains the output at the value which was present at the time of
echo loss. The “Min - Max Mode” (Switch 3 On) drives the output to either the
minimum value (0V or 4mA or the maximum value (10V or 20mA) when the echo is
lost. Minimum or maximum value is selected by Switch 4.

Switch 4:

Min - Max Default

On* = Default to maximum output value at loss of echo
Off = Default to minimum output value at loss of echo

Switch 4 selects the output response to loss of echo when “Min - Max Mode” is
selected by Switch 3 (see above).

Response Speed Adjustment

The speed of the output response is set using the single-turn potentiometer (see
Figures 1 and 6). There are six values for response speed, which relate directly to the
number of sensing cycles over which the output value is averaged. The response
value is set by aligning the slot of the potentiometer with one of the marked positions.
The positions and their corresponding settings are identified in Figure 6.

Inverse Slope
Switch 1 Off

Direct Slope
Switch 1 On

Near

Window

Limit

Far

Window

Limit

10V dc

(20mA)

0V dc

(4mA)

Figure 5. Output Slope

Res

Position

Response Speed

1
2
3
4
5
6

40 milliseconds (2 cycles)
80 milliseconds (4 cycles)

160 milliseconds (8 cycles)

320 milliseconds (16 cycles)
640 milliseconds (32 cycles)

1280 milliseconds (64 cycles)

Figure 6. Response adjustment positions and their corresponding response speed values

1

2

3

4

5

6

NOTE: This example shows the potentiometer

set at position number 4. There are no
numbers on the actual product label.

+

Response Speed Settings