The following table lists the error messages and their associated meanings:

Error

ASSOCIATED MEANING

TYPICAL CAUSE

1

Lamp failure

Lamp has too low an output for proper

turbidity measurement.

2

General Calibration failure:

There is not sufficient signal

between the 0.02 NTU and 10.0

NTU standards used for

calibration

Either the wrong standards were used in

calibration or there is an internal sensor

failure.

3

General Calibration failure:

There is not sufficient signal

between the 10.0 NTU and 100.0

NTU standards used for

calibration

Either the wrong standards were used in

calibration or there is an internal sensor

failure.

4

General Calibration failure:

There is not sufficient signal

between the 100.0 NTU and

1000.0 NTU standards used for

calibration

Either the wrong standards were used in

calibration or there is an internal sensor

failure.

Err

Multiple error conditions are met