Kenco Engineering SmartSonic Transmitter (Integral) User Manual

Page 7

Advertising
background image

Page 7

NOTE: Within some programs, the digits in the display can be individually modified. This is indicated by a flashing digit.
In this case the

ENT and ESC buttons are used to move between digits. The BACK and NEXT buttons are used to

change the value of the flashing digit.


INITIAL SET-UP PROGRAM
Prior to calibration, you must verify that the SmartSonic is set for your application. There are five settings to consider.
These settings can be accessed through the

Pr.02 Program.


Initial Set-up Program Menu

Pr.10

Pr.24

Pr.23

Pr.22

Pr.21

Pr.20

Factory

Default

Units

(Level)

Units

(Flow)

Flume

Type

Measurement

Range

Application

Type

Pr.01

Reset


Here are the settings to consider:

Application Type

Pr.24

Measurement Range

Pr.23

Flume Type (Open Channel Flow only)

Pr.22

Units (Flow) (Open Channel Flow only)

Pr.21

Units (Level)

Pr.20


The following pages will take you through this process. Start with

Pr.24. If you select Level (L), Pr.22 and Pr.21 will not

be available.