Problem

Cause

Action

Erratic LCD
display

Samples have low ionic
strength (lack salt);
e.g. distilled, deionized,
boiled, lake water (high
pressure)

For each 50 mL of sample add
1 drop (50 mL) of saturated
Potassium Chloride (KCl). No
alteration in pH will occur by
inert KCl.

Unexpected
readings

Buffers

Calibrate with fresh buffers

Buffers

Calibrate with buffers that
bracket sample pH

Display frozen

HOLD function

Press MODE/HOLD or turn
meter off

Button response

Remove batteries press ON/OFF
button for 3 seconds. Reinsert
batteries (stored data will be
lost)

Steady “-1 ”
display

Wait

Reading not stable yet