Blank done, 09 ppm – LaMotte SMART 2 Electronic Soil Lab User Manual
Page 35
The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.
002 Aluminum
Blank Done
* Scan Sample
Place the reacted sample in the chamber, close the
lid and press
Q/ENTER to scan sample. The result
will be displayed.
002 Aluminum
* Scan Sample
The displayed result can now be standardized. Use
the
s or t buttons to scroll to Calibrate. Press
Q/ENTER to select.
002 Aluminum
0.09 ppm
* Scan Sample
Next Test
Previous Test
%T/Abs
Calibrate
Scan Blank
A reverse font (dark background with light
characters) will appear to indicate that the reading
can be adjusted. Use
s or t to scroll to the
concentration of the sample, 0.10 ppm in this
example.
002 Alumninum
0.09
* Calibrate
Set the calibration by pressing
Q/ENTER to select
Calibrate
.
002 Aluminum
0.10
* Calibrate
SMART2 SOIL MANUAL 1.07