LaMotte HESTER SOIL TEST ST-H User Manual

Page 2

Advertising
background image

PRO CE DURE

1.

Fill a test tube (0204) approximately one-third full of soil. Use the
Model PWB-1 Demineralizer Bottle (1155) to add demineralized
water to the tube, until it is one-half inch from the top. Cap and
shake until the soil is well dispersed.

2.

Add 5 drops of Soil Flocculating Reagent (5643). Cap and shake to
mix. Allow contents to settle before proceeding to Step 3.

3.

Use a 1 mL pipet (0354) to transfer 1 mL of the clear solution above
the soil to one of the large depressions on a spot plate (0159).
Transfer a second 1 mL sample to the other large depression on the
spot plate.

4.

To the first sample on the spot plate, add two drops of *Duplex
Indicator (2221). Compare the resulting color reaction against the
Duplex Color Chart (1313). NOTE: The wide range pH test result
indicates which narrow range indicator and color chart should be
selected to perform a more precise pH test. Choose the narrow range
indicator and appropriate chart with a mid-point that is as close as
possible to the value obtained in the wide range test.

INDICATOR

pH
RANGE

INDICATOR
CODE

COLOR
CHART CODE

Bromcresol
Green

3.8-5.4

2207

1328

Chlorphenol Red 5.2-6.8

2209

1329

Bromthymol Blue 6.0-7.6

2210

1331

Phenol Red

6.8-8.4

2211

1332

EXAMPLE:

If the wide range test result is pH 6.0, choose the Chlorphenol

Red Indicator (2209) & the Chlorphenol Red Color Chart (1329) for
Step 5.

(over)

Advertising