LaMotte Water Analysis Kit for Home Brewers User Manual

Page 10

Advertising
background image

pH

QUANTITY

CONTENTS

CODE

1

pH Tracer PockeTester

1741

10

pH 7.0 Mini Buffer Tablets

3894A

10

pH 4.0 Mini Buffer Tablets

3983A

Sample should be below 122 °F (50 °C). Let cool if necessary.
NOTE: The purpose of ATC in this meter is to maintain electrode calibration

across the capability range of 32-194 °F (0-90 °C). It is recommended that

wort samples be cooled to room temperature (68 °F/20 °C) before testing.

This allows better comparison of measurements at different temperatures

due to the natural change in wort pH due to changes in temperature, and

improves electrode life.
1. Remove cap. Press ON/OFF to turn meter on.
2. Dip sensor in sample water.
3. Record pH after reading becomes stable and display stops fl ashing.
4. Press MODE to hold reading and store reading in memory
5. Press ON/OFF to shut meter off.
6. Store in pH 4.0 buffer. Store upright.

Daily Calibration Procedure
1. Fill tube (0715) to 20 mL line with distilled water. Pour into small container.

Add pH 7.0 Mini Buffer Tablet (3984A). Stir until tablet disintegrates.

2. Remove cap. Press ON/OFF to turn the meter on.
3. Dip sensor in buffer.
4. Press CAL. END will be displayed and the circled 7 indicator will be

displayed.

5. Repeat Steps 1–4 with pH 4 Mini Buffer Tablet (3983A).

Save pH 4 Buffer

to use as storage solution. Tester is calibrated.

Before First Use or After Storage
1. Hold meter by top battery compartment. Swiftly tap back of meter

downward into palm - not a hard surface - to move electrolyte into

electrode tip.

2. Remove cap. Soak electrode in pH 4.0 buffer or tap water for 10 minutes.