Safeguards and precautions – Monarch Instrument EXAMINER 1000 User Manual

Page 2

Advertising
background image

Safeguards and Precautions

1. Read and follow all instructions in this manual carefully, and retain

this manual for future reference.

2. Do not use this instrument in any manner inconsistent with these

operating instructions or under any conditions that exceed the
environmental specifications stated.

3. This instrument is not user serviceable. For technical assistance,

contact the sales organization from which you purchased the
product or Monarch Instrument directly.

In order to comply with EU Directive 2002/96/EC on Waste Electrical and
Electronic Equipment (WEEE):
This product may contain material which
could be hazardous to human health and the environment. DO NOT
DISPOSE of this product as unsorted municipal waste. This product needs
to be RECYCLED in accordance with local regulations, contact your local
authorities for more information. This product may be returnable to your
distributor for recycling - contact the distributor for details.

Monarch Instrument’s Limited Warranty applies. See www.monarchinstrument.com
for details.

Warranty Registration and Extended Warranty coverage available online at
www.monarchinstrument.com.

EXAMINER 1000 SPECIFICATIONS

Vibration Sensor:

Piezoelectric accelerometer 100 mV/g with magnetic base,
probe and 5 foot cable to BNC

Display:

LCD 3.5 digit , measurement, hold, low battery indicator

Measurement

Acceleration 0.01-19.99 g (RMS)

Range:

Velocity 0.01-19.99 in/sec., 0.1-199.9 mm/sec (RMS)
Envelope 0.01-19.99 ge (peak)

Frequency Range:

Overall 10 Hz - 10 kHz
Envelope 0.5 kHz-10 kHz

Output:

Sensor Excitation: 18 Vdc @ 2 mA (BNC)
Audio Out: 3.5 mm mini plug; 250 mW into 8 ohms, 150 mW
into 32 ohms; Adjustable volume control with off position

Power:

(2) “AA” cells

Operating Time:

20 hours continuous without headphones

Weight:

Instrument: 7 oz. (0.19 kg); Complete Kit: 2.85 lb (1.30 kg)

Dimensions:

6.3 x 3.3 x 1.25” (152 x 83 x 32 mm)

Operating conditions:

-14° to 122°F (-10° to 50°C)

Use the chart below as a guide to judge the overall vibration severity of your equipment.
Refer to page 12 for further details.

VIBRATION SEVERITY PER ISO 10816-1

Vibration Velocity Vrms

Machine

Class I

Class II

Class III

Class IV

small

medium

large rigid

large soft

in/s mm/s

machines

machines

foundation

foundation

0.01 0.28
0.02 0.45
0.03 0.71
0.04 1.12
0.07 1.80
0.11 2.80
0.18 4.50
0.28 7.10
0.44 11.2
0.71 18.0
1.10 28.0
1.77 45.0

good

satisfactory

unsatisfactory

unacceptable