Electrical diagram (bottom view) – Amico Ceiling Column Motorized Retractable User Manual

Page 7

Advertising
background image

*Please Select One of the Following:

Amico recommended circuit schedule based on layout

Custom circuit schedule. Please specify on circuit chart

*Please specify electrical details:

Type of wire:

THHN

XHHW

RW90

AWG Wire Gauge:

Wire Color:

10 & 12 LIVE/HOT

10 & 12 NEUTRAL

10 GROUND


Isolated

Power

Please select one of the following

All receptacles shall be mounted with

ground

pin

“DOWN”

All receptacles shall be mounted with

ground pin “UP” (Default)

Receptacle Type:

US/CAN

*BRITISH

(simplex only)

*GERMAN

(simplex only)

Circuit Chart

Electrical Diagram (Bottom View)

NOTE:

1.

Two electrical/communication/medical gas devices

cannot be placed together at any corner.

2.

If type of wire & gauge is not specified, Amico

will use the following:

For US/International

LIVE/HOT - 12 Ga. THHN Wire Black

NEUTRAL - 12 Ga. THHN Wire White

GROUND - 10 Ga. THHN Wire Green

For Canada

LIVE/HOT - 12 Ga. RW90 Wire Black

NEUTRAL - 12 Ga. RW90 Wire White

GROUND - 10 Ga. RW90 Wire Green

3.

Conduit: 9/16” [14.28mm] fl exible metal conduit.

4.

All conduits and wires will extend a minimum of12”

[304.8mm] above fi nish column for connection

to riser plate.

5.

A pull-string will be provided to be connected by

others for all ground jacks and data provision.

6.

British or German Receptacle cannot be positioned

side by side ie: A1 and A2.

Approval Signature

Date

Phone No.

www.amico.com

7

12 gauge

12 gauge

Black

White

10 gauge

Green