EMC-Inspection Report

Immunity test

standard

IEC 60601 test

level

Compliance level

Electromagnetic environment -

guidance

Voltage dips, short

interruptions and voltage

variations on power

supply lines
IEC 61000-4-11

<5% UT (0,5 cycle)

40% UT (5 cycles)

70% UT (25 cycles)

<5% UT for 5s

<5% UT (0,5 cycle)

40% UT (5 cycles)

70% UT (25 cycles)

<5% UT for 5s

Mains power quality should be that of a typical

commercial or hospital environment.

Note: UT is the a.c. mains voltage prior to application of the test level.

Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8

3 A/m

100 A/m

Power frequency magnetic fields should be at

levels characteristic of a typical location in a typical

commercial or hospital environment.

Guidance and manufacturer’s declaration – electromagnetic immunity

The Gamma 50 LED is intended for use in the electromagnetic environment specified below. The customer or the user of the

Gamma 50 LED should assure that it is used in such an environment.

Electromagnetic environment - guidance

Portable and mobile RF communications equipment should be used no closer to any part of the Gamma 50 LED, including cables,

than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Immunity test standard

IEC 60601 test level

Compliance level

Recommended separation distance C

Conducted RF

IEC 61000-4-6

3 Vrms

150 kHz to 80 MHz

V1 = 10 Vrms

(estimated value)

150 kHz to 80 MHz

d = 0.35 √P 150 kHz to 80 MHz

Radiated RF

IEC 61000-4-3

3 V/m

80 MHz to 800 MHz

E1 = 10 V/m

80 MHz to 800 MHz

d = 0.35 √P 80 MHz to 800 MHz

Radiated RF

IEC 61000-4-3

3 V/m

800 MHz to 2,5 GHz

E2 = 10 V/m

800 MHz to 2,7 GHz

d = 0.7 √P 800 MHz to 2,7 GHz

Table 3: Immunity (All Devices)

Table 4: Immunity (Not Life-Supporting Devices)