Melt Calibration

Before Precision Melt Analysis software can analyze data generated on
a CFX96, CFX96 Touch, CFX96 Touch Deep Well, CFX Connect, CFX384,
or CFX384 Touch real-time PCR detection system, a melt calibration must
be performed. The procedure requires running an experiment and then
importing the data into Precision Melt Analysis software to generate a melt
calibration file. The software is shipped with a calibration kit that includes
the following materials:

n

Melt calibration DNA standard

n

Melt calibration primers

n

Precision melt supermix (catalog #172-5110)

Additional Materials Required

In addition to the components provided in the melt calibration kit,
the following materials are required:

n

PCR-grade tubes

n

Nuclease-free water

n



Microseal

®

'B' adhesive seals, optically clear (catalog #MSB-1001)

n



Hard-Shell

®

thin-wall 384-well skirted PCR plates with clear shell

and white wells (catalog #HSP-3805) for use with a CFX384 or
CFX384 Touch system

n

Hard-Shell thin-wall 96-well skirted PCR plates with white shell and
clear wells (catalog #HSP-9601) or white wells (catalog #HSP-9655)
for use with a CFX96, CFX96 Touch, CFX96 Touch Deep Well, or
CFX Connect system

Preparing a Melt Calibration Plate

1. Add the required volume of each component to an appropriately

sized tube (Table 1).

2. Cap the tube and gently mix the reaction components by vortexing.

3. Briefly centrifuge the tube to remove air bubbles and collect

contents at the bottom of the tube.

4. Add the appropriate volume of the mixture into each well of a

reaction plate:

n



For a CFX96, CFX96 Touch, CFX96 Touch Deep Well, or
CFX Connect system, add 20 µl of the reaction mixture to each
well of a 96-well plate

n

For a CFX384 or CFX384 Touch system, add 10 µl of the reaction
mixture to each well of a 384-well plate

5. Seal the reaction plate with a Microseal 'B' adhesive seal. Centrifuge

the plate at 1000 x g for 2 min to move all the reaction components
to the bottom of the wells.

Performing a Melt Calibration Experiment

A melt calibration experiment is run on a CFX96, CFX96 Touch,
CFX96 Touch Deep Well, CFX Connect, CFX384, or CFX384 Touch real-
time PCR detection system and analyzed using CFX Manager

software.

To run a melt calibration to generate a melt calibration data file:

1. Turn on the CFX system.

2. Double click the CFX Manager software desktop icon to launch

the software.

3. Select

User-defined from the Run setup tab of the Startup Wizard

(Figure 3) to launch the Run Setup window.

Table 1. Reaction setup for a melt calibration plate.

Volume for

Volume for

Component

CFX96 System

CFX384 System

Precision melt

1,200 µl

2,250 µl

supermix
Melt calibration

120 µl

450 µl

DNA standard
Melt calibration

14.4 µl

27 µl

primers
Nuclease-free water 1,065.6 µl

1,773 µl

Total

2,400 µl

4,500 µl

Table 2. Melt calibration PCR protocol.

Number

Cycling Step

Temperature Time

of Cycles

Enzyme activation

98°C

2 min

1

Denaturation

98°C

5 sec

35

Annealing/extension 55°C

10 sec

95°C

1 min

1

70°C

1 min

1

Melt curve

70–95°C

10 sec/step 1

(in 0.2°C

increments)

Fig. 3. Startup Wizard in CFX Manager software.