Troubleshooting – Bio-Rad S3™ Cell Sorter User Manual

Page 31

Advertising
background image

Instruction

Manual | 21

Troubleshooting

5

Table 7 describes possible errors and causes with troubleshooting steps.

Table 7. Troubleshooting table.
Error

Possible Causes

Troubleshooting Steps

Biosafety system icon

System not configured to have

Contact technical support to correct the

does not show up in

a biosafety system

configuration of the system

ProSort

software

There is no fan noise

System is very quiet

Stand closer to fan unit and listen for fan

coming from the unit

noise. There should be a quiet hum

Not plugged in correctly

Check AC power cord for a tight connection
on system and in the wall

HEPA may be full and system

Check the biosafety system status in ProSort

has shut down.

software to see if the system’s HEPA is at
less than 10% capacity. HEPA filter may need
replacement

There is a rattling noise

Loose fitting or latch

Check all connections and latches. The

coming from the air

system should have a quiet hum and low

generator fan unit

vibration

There is a large gap

Misaligned magnetic attachment

Re-attach the access door. Adjust by gently

between magnetic

bending the access door to eliminate the

connections on

gap. If the gap remains and the access door

access door

border matches with the panel borders, then
system is functional

The panels do not stay

Panels do not lie well against air

Roll the panels slightly tighter so they can lie

up when opened

generator and plastic top panel

evenly on the plastic top panel

The access doors are

S3 or S3e cell sorter is in the

Adjust the S3 or S3e cell sorter carefully to

not aligned with the

improper location

line up with the access doors as much as

system components

Improper location may not maintain safety