3. Adjust the screw that attaches the camera to the

bracket and the screw that attaches the bracket
to the hood until the camera stands upright and
appears to be centered. You can check the horizontal
alignment of the camera by looking at the side of the
instrument and the vertical alignment by looking up at
the filter holder inside the hood.


Connect the Cables

Note: DO NOT plug the FireWire cable into the
computer at this time.

Connect all of the cables to the camera, the hood,
and the computer. Each system will have the following
connections:

Camera

1. Power supply
2. FireWire cable

Universal hood

1. Serial cable
2. Motorized zoom lens (MZL) cable (from camera lens)
3. Power cable

Computer

1. Serial cable from universal hood II to COM 1 port (PC)
2. Serial cable from universal hood II to USB via USB

adaptor (Mac)

Locate the FireWire Connection
on the Computer

PC:

Some PCs may require the addition of a PCI FireWire
(IEEE 1394) adaptor card.

1. Make sure that the computer is powered off and

the power cord is disconnected from the PC while
installing the FireWire adaptor card.

2. Power on the computer.

3. The Windows 2000 and Windows XP operating

systems will recognize the card and automatically
install the appropriate driver for it.

Mac:

Current Mac systems have onboard FireWire ports.

Install Quantity One Software

PC:

1. Insert the software CD and follow the default

installation instructions on your screen.

2. To install the appropriate camera drivers, select the

following on your screen:

3. Follow the default installation routine to complete the

installation of the software.

4. After installation is complete, plug the FireWire cable

from the camera into the PC. Ensure that the power
supply for the camera has power connected to it.
Additionally, on the ChemiDoc XRS, make sure that
the power switch is turned on.

Horizontal alignment

Vertical alignment

ChemiDoc XRS
assembly

Center lens in the
opening of the hood
(as viewed from inside
the hood)

Gel Doc XR assembly

Select “Will be installed

on local hard drive.”

Select the Gel Doc XR or ChemiDoc

XRS option, depending on your

system type

Power cable from power supply to camera

MZL cable

from lens to

universal hood

Cabinet controls from universal hood to serial port on computer

FireWire cable from camera

to FireWire card or port