5. Reconstitute a single vial of standard in 781 µl of a diluent similar to your

final sample type or matrix. Reconstitute the quality controls in 250 µl of
the same diluent, as shown below. Vortex for 5 sec and incubate all vials
at once on ice for 30 min. (Section 4)

Note:

The quality controls are ready to use after reconstitution. No dilution is

needed. Only premixed panels and x-Plex kits include controls.


6. Prepare a fourfold standard dilution series and blank as shown below.

Vortex for 5 sec between liquid transfers. (Section 4)

7. After thawing samples, prepare as shown below. (Section 5)

8. Vortex coupled beads for 30 sec and dilute to 1x in Bio-Plex assay buffer

as shown below. Protect from light. (Section 6)

# of Wells

20x Beads, µl

Assay Buffer, µl

Total Volume, µl

96

288

5,472

5,760

150 150 150 150 150 150 150 150

S1

S2 S3 S4 S5 S6 S7 S8

Blank

Diluent, µl

50

50 50 50 50 50 50

Transfer Volume, µl

Reconstituted

Standard

Sample Type

Diluent for Standard and Controls*

Add BSA

Serum and plasma

Standard diluent HB

None

Culture media, with serum

Culture media

None

Culture media, serum-free

Culture media

To 0.5% final

Lavage, lysate, other fluids

Sample diluent HB

To 0.5% final

* If using diluents other than standard diluent HB, users must establish their own control ranges.

Sample Type

Diluent

Add BSA

Sample Dilution

Serum and plasma

Sample diluent HB

None

Fourfold (1:4)

Culture media, with serum

Culture media

None

Neat to 1:10

Culture media, serum-free

Culture media

To 0.5% final

Neat to 1:10

Lavage, lysate, other fluids

Sample diluent HB

To 0.5% final

User optimized

Bio-Plex Pro Assay Quick Guide 1