Bio-Rad Bio-Plex Pro™ Rat Cytokine, Chemokine, and Growth Factor Assays User Manual

Bio-plex pro, Assays, Quick guide 4

Advertising
background image

This guide can be used to prepare and run a full 1 x 96-well assay plate.
Refer to the complete instruction manual for more information on a given
step. New users can download the manual, which includes detailed
instructions and a list of kit components, at www.bio-rad.com/bio-plex.

Initial Preparation

1. Plan the plate layout.

2. Start up/warm up the Bio-Plex

®

system (30 min).

n

Bring assay buffer, wash buffer, and sample diluent to room

temperature (RT). Keep other items on ice until needed

n

Begin to thaw frozen samples

3. Prime wash station for flat bottom plate or set vacuum manifold to

–1 to –3" Hg for filter plate.

4. Calibrate the Bio-Plex system by following the prompts within

Bio-Plex Manager

software. This can be done now or during an

assay incubation step.

5. Reconstitute a single vial of standards in 500 µl of a diluent similar to the

final sample type or matrix. Vortex for 5 sec and incubate on ice for 30 min.

Bio-Plex Pro

Assays

Quick Guide 4

IMPORTANT!

Pay close attention to vortexing, shaking, and incubation instructions.

Deviation from the protocol may result in low assay signal and assay variability.


For use with

Instruction Manual #

Human, Mouse, and Rat Cytokine Assays

10014905

Sample Type

Diluent for Standards

Add BSA

Serum and plasma

Standard diluent

None

Culture media, with serum

Culture media

None

Culture media, serum-free

Culture media

To 0.5% final (w/v)

Advertising