Calibrating the Fraction Collector X–Y Arm

To calibrate the Fraction Collector X–Y arm

Caution:

Using this function overwrites the previous calibration.

1.

Remove the drop former top by twisting it counterclockwise and gently lifting it
out.

2.

Press X–Y Axis and then Next.

3.

While looking through the drophead, center it over tube A15 using the arrow
keys, and press Next.

4.

Center the drophead over tube A1 using the arrow keys, press Next.

5.

Center the drophead over tube B90 using the arrow keys, press Save.

6.

Replace the top of the drop former by inserting it into the drophead and twisting
it clockwise.