Troubleshooting Guide

Problem

Possible Cause

Possible Solution

Low protein

Low levels of protein

Increase load on column

concentration

in sample

Concentrate sample in
SpeedVac

High protein

High levels of protein

Decrease the load

concentration

in sample

applied to column

Buffer does not flow

Frit is clogged with

Prefilter sample prior

through column

sample debris

to loading

Protein does not bind

Incorrect column chosen

CEX column is for

to column

basic proteins; AEX
column is for acidic
proteins

Ionic detergents can

Avoid using detergents

prevent sample from

with opposing net

binding

charges to that of ion
exchange column.
Detergents can be
removed using
Bio-Beads

®

SM-2

macroporous beads
(catalog #152-3920)

Ionic strength too high

Dilute sample or desalt

pH not correct

Check pH of loading
buffer

Protein does not elute

Salt concentration too

Increase salt

from column

low in elution buffer

concentration to 2 M
and/or change pH

Sample is dilute after

Excess packing buffer

Remove excess packing

centrifugation

was not removed

buffer by centrifugation
prior to loading sample