Aurum, Serum protein mini kit, Protocol overview – Bio-Rad Aurum Serum Protein Mini Kit User Manual

Page 9

Advertising
background image

7

Fig. 2. Aurum serum protein mini kit protocol overview

Aurum

Serum

Protein Mini Kit

Protocol Overview

Column Setup

1.

Place serum protein column in a test tube

for 5 min to allow resin to settle.

2.

Remove cap and break tip from column

and return to test tube to start gravity

flow in column.

3.

Wash column with 1 ml of serum protein

binding buffer using gravity flow. Repeat.

4.

Place column in empty 2.0 ml collection tube

and centrifuge for 20 sec at 10,000 x g to

dry resin bed. Discard collection tube.

5.

Place a yellow column tip on bottom of column

and place into a clean 2.0 ml collection tube

labeled “unbound”.

Sample Binding and Purification

6.

In a separate tube, prepare sample by diluting

60 µl serum or plasma with 180 µl of

serum protein binding buffer.

7.

Add 200 µl of diluted serum to top of

resin bed in column.

8.

Gently vortex column and repeat after

5 and 10 min. Allow column to sit

an additional 5 min.

Collection of Purified Samples

9.

Remove yellow tip from column and return

to tube. Centrifuge column for 20 sec at

10,000 x g, collecting protein fraction in

“unbound” collection tube.

10.

Using same collection tube, wash column

with 200 µl of serum protein binding buffer.

11.

Centrifuge column for 20 sec at 10,000 x g,

collecting protein fraction in same “unbound”

collection tube. Discard serum protein column.

12.

The combined fractions contain the albumin-

and IgG-depleted serum or plasma sample.

The sample is now ready for gel analysis.

1 ml
binding
buffer

60 µl
serum

200 µl
diluted
serum

200 µl
binding
buffer

180 µl
binding
buffer

Vortex

2x

3x

4110021B.qxd 8/14/2003 2:51 PM Page 7