2 wiring, 3 setup procedure, 4 specifications – Cooper Instruments & Systems DFI 1650 Multi-Channel Digital Force Indicator User Manual

Page 67: 5 channel menu, 1 diagnostics sub-menu

Advertising
background image

CF 66

61

Rev. C 2/05


In the SYSTEM->CONFIGURATION sub-menu, channels of this type are identified as LIMIT 13-16.

13.2 Wiring

The pin-out for the Relay Output channel’s 12-pin connector is shown in the following table.

Table 13-1: Relay Output Channel Pin Connections

Pin Label

Function

Reference

Pin

1 (top)

RELAY1 NC

Relay 1 normally closed

2

2

RELAY1 C

Relay 1 common

-

3

RELAY1 NO

Relay 1 normally open

2

4

RELAY2 NC

Relay 2 normally closed

5

5

RELAY2 C

Relay 2 common

-

6

RELAY2 NO

Relay 2 normally open

5

7

RELAY3 NC

Relay 3 normally closed

8

8

RELAY3 C

Relay 3 common

-

9

RELAY3 NO

Relay 3 normally open

8

10

RELAY4 NC

Relay 4 normally closed

11

11

RELAY4 C

Relay 4 common

-

12 (bottom)

RELAY4 NO

Relay 4 normally open

11

13.3 Setup Procedure

If you are not familiar with operating the instrument in the SETUP menu mode, see “SETUP Menu mode” in Section
3.5 A listing of all menu items is given in “Setup Menu Reference” in Chapter 18.

Step 1: Wire to the channel’s connector.
See the “Wiring” section earlier in this chapter for details.

Step 2: Use the appropriate Limit Menu to setup the limits.
See “Limits”, Chapter 9, for information about limits.

A SensoCode program running on a Mathematics Virtual Channel may override the default behavior of the relay

outputs. Consult the System Calibration Sheet included with your instrument for details.

13.4 Specifications

Relay Output

Quantity and Type

4 form C

Contact Rating

0.5A @ 50VAC max. (consult factory for 125VAC max. operation)
1A @ 30VDC max.

13.5 Channel Menu

The Relay Output channel’s SETUP menu allows you to manually turn the relays on and off to verify that they are
functioning. To change the operation of the limits, see “Limit Menus” in Section 9.3.

Detailed instructions on operating the instrument in the SETUP Menu mode can be found in “SETUP Menu mode”.
A diagram of all menus is located in “Setup Menu Reference” in Chapter 18.

13.5.1 DIAGNOSTICS Sub-Menu

RELAY 1 Menu Item

The options for this menu item are:

• “ON” will turn the relay on.

• “OFF” will turn the relay off.

RELAY 2 Menu Item

The options for this menu item are:

• “ON” will turn the relay on.

Advertising