Cooper Instruments & Systems Series M2 Digital Force Gage User Manual

Page 9

Advertising
background image

Series 2 Digital Force Gauges

User’s Guide


8

7 DIGITAL FILTERS


Digital filters are provided to help smooth out the readings in situations where there is mechanical
interference in the work area or test sample. These filters utilize the moving average technique in which
consecutive readings are pushed through a buffer and the displayed reading is the average of the buffer
contents. By varying the length of the buffer, a variable smoothing effect can be achieved. The selection
of 1 will disable the filter since the average of a single value is the value itself.

To access digital filter settings, select Filters from the menu. The display will appear as follows:


Two filters are available:

Current Reading – Applies to the peak capture rate of the instrument.

Displayed Reading – Applies to the primary reading on the display.


Available settings: 1,2,4,8,16,32,64,128,256,512,1024. It is recommended to keep the current reading
filter at its lowest value for best performance, and the displayed reading filter at its highest value for best
stability.

8 CALIBRATION

8.1 Initial Physical Setup

The gauge should be mounted vertically to a test stand or fixture rugged enough to withstand a load
equal to the full capacity of the instrument. Certified deadweights or master load cells should be used,
along with appropriate mounting brackets and fixtures. Caution should be taken while handling such
equipment.

8.2 Calibration Procedure

1. Select

Calibration from the menu. The display will appear as follows:

2. Press

DIRECTION to invert the display, if desired. ENTER to continue. The display will appear as

follows:

DIGITAL FILTERS

(1 = Fastest)

Current Reading

8

Displayed Reading
1024

CALIBRATION

To invert the
display, press the
DIRECTION key.

THEN PRESS ENTER

Advertising