0 available options for all instruments, 0 specifications, 0 setup – Cooper Instruments & Systems ACX-1 Aircraft Cable Tension Meter User Manual

Page 6

Advertising
background image

CF 155

6

9/25/07

5.1 STANDARD INSTRUMENTS

MODEL RANGE RESOLUTION STANDARD

CABLE

SELECTIONS

ACX-250

20-250 LBS

0.5 LB

1/16", 3/32", 1/8", 5/32"

ACX-500

40-500 LBS

1.0 LB

3/32" 1/8", 5/32"

ACX-CWT

20-250 LBS

0.5 LB

CWT-5, CWT-10, CWT-15, CWT-20, CWT-25

ACX-CMB

20-250 LBS

0.5 LB

1/16", 3/32", 1/8", 5/32", CWT-5 thru CWT-25


6.0 AVAILABLE OPTIONS FOR ALL INSTRUMENTS:

Note: Add option designator to the end of the part number. (For example, a Model ACX-250-E denotes a standard,
ACX-250 with the optional RS-232, serial output function installed).

• -A

Analog Output option.

0-5 VDC or 4-20 mA with Software-definable ending sequences. Provided with
10’ sequences. Provided with 10’ cable. Data outputted at 40 Hz.

• -E

RS-233 Serial Output option.

Provided with 10’ cable to Interface with your receiving device. Select data
sampling rate from 1, 2 or 5 Hz.

Input Real-Time Serial Data into any Windows® application using the optional
software WinWedge Pro® for Windows, or configure your Window’s® operating
system to capture the data via Hyper Terminal®.

7.0 SPECIFICATIONS

• Approximate weight is 1-1/2 lbs (Weight varies depending upon instrument configuration).

• Re-chargeable NIMH battery with power supply provided. Approximately 6 hours of operation per battery charge

depending upon backlighting intensity.

• Power Supply operates with input voltages from 100 – 240V and includes several, interchangeable “blades”

allowing use with European, US, Australian, and other plug configurations.

• Automatic shutoff after 10 minutes of non-use.

• Instruments may be operated continuously while connected to power supply.

• Durable, lightweight carrying case with protective foam inserts.

• CE Certification complying with heavy, industrial immunity standards.

8.0 SETUP

CHECKING ACCURACY. If you want to verify the accuracy of your instrument, simulate a tension load on your
cable by suspending known weights to a sample length of the same type of cable. When you engage the meter this
weight value should display. Thus when a 100 lb weight is freely suspended from the cable, the tension will be 100
lbs, and your instrument should indicate 100 lbs.

8.1 Custom Calibrations for the ACX-Series of Digital Tension Meters

Warning: The Custom Calibration feature is password protected. In order to perform a Custom Calibration a
series of key inputs are required. If you enter the password sequence and the press Enter while an existing
calibration in the menu is selected/highlighted, it will void that current calibration selection.

8.2 Performing a 5-Point, Multilinerization Calibration

Step 1: From the Main Display press

or

arrows until you reach Setup. Press ENTER. Press the

arrow button

until you reach Cal Tension. Press ENTER. The screen will display: Setup, Calibrate Tension, Enter Password.

Advertising