COVIDIEN Kangaroo™ Joey User Manual

Kangaroo joey, Loading & rinsing of feeding pump, Enteral feeding & flush pump

Advertising
background image

Kangaroo Joey

Enteral Feeding & Flush Pump

Loading & Rinsing of Feeding Pump

Do not use feeding sets for longer than 24 hours.

Proper loading & unloading of a feeding set

Proper rinsing of the feeding set

1) Feed & flush valve assembly, 2) Feed only valve

assembly, 3) Distal end of feeding set, 4) Pump

1) Grasp thumb tab on valve and insert into left

pocket, ensure valve is fully seated. Thumb tab

should align with slot in blue door.

2) Grasp black ring retainer, gently wrap tubing

around rotor, and insert retainer directly into
right pocket.

Do not overstretch tubing.

1) Disconnect feeding set from patient before rins-

ing. The feeding set must remain loaded during

the rinsing procedure.

2)

CAUTION-

if set is removed from the pump,

it may damage the set and prevent use and will

negativley effect the system accuracy.

1

2

3

4

3) To properly unload a pump set, power the

pump off and wait 10 seconds before removing

the feeding set.

3) Empty contents of feed bag.

4) Fill the feed bag with water to approximatley

half of the total capacity. Properly seal the bag

and shake for 5 to 10 seconds.

Advertising