SYMPTOM

PROBABLE CAUSE

CORRECTIONS

Hold Error

Pump in HOLDING mode for
more than 10 minutes.

Press

}

CONTINUE to return to HOLDING screen, or

press and hold the Power button to turn off. Several options
are available from the HOLDING screen, such as RUN or
ADJUST SETTINGS.

Flow Error

Occlusion in pump-patient line.
(downstream occlusion)

Check for and correct the occlusion in the line between pump
and patient. Replace Pump Set if error cannot be resolved.

Feed Error

Empty bag or occlusion in
bag-pump line. (upstream
occlusion)

Check for empty bag and refill, or check for and correct
the occlusion in the line between the pump and feed bag.
Replace Pump Set if error cannot be resolved.

Flow Error
Use > 24 Hrs

Pump Set in use beyond
recommended length of time.

Informational warning message blinks on upper left of screen.
It is recommended that the Pump Set be replaced.

Pump Set
Dislodged

Pump Set not properly loaded
to pump.

Check black ring retainer (MISTIC) and valve; reload Pump
Set and restart pump to continue. Replace Pump Set if error
cannot be resolved.

Battery Low

Battery charge too low.

Immediately plug pump into A/C outlet to recharge battery
and to provide power for pump.

Rotor Error

Pump Set tubing not properly
loaded on rotor, or some
other unusual rotor operating
condition.

Check that the tubing is loaded on the rotor and that it is not
damaged, torn, etc. Reload tubing or replace Pump Set.

System Error

General error caused by many
factors. See manual for error list.

Power down pump and attempt restart to clear error. Call
customer service with screen code if error cannot be resolved.

Menu Selections

Programming & Menu

Selection Buttons

Status LEDs
Red=Error
Yellow=Hold/Pause
Green=Normal Operation

Power Button

15 Hampshire Street

Mansfield, MA

02048

1-800-962-9888

508-261-8000

www.covidien.com

COVIDIEN, COVIDIEN with logo, Covidien logo,

positive results for life and Kangaroo

are U.S. and/or internationally registered trademarks of Covidien AG.

©2010 Covidien.

H6206-10M-0610

Listed below are some of the probable causes of alarm conditions and their corrections.

Kangaroo

Joey Enteral Feeding Pump

Troubleshooting Guide