that have flat and concave surfaces “on edge” will enable the

condensate to drain off and more readily dry, usually without

spotting. An additional cause of spotting can be traced to the

instrument wraps. During laundering procedures, it is vital that

detergents are thoroughly rinsed out, and that the final rinse is

prepared so that the wraps have a pH between 6.8 and 7.0. In

addition, healthcare professionals should check the cleanliness

of the sterilizer chamber. Steam can lift soil and poorly rinsed

chamber cleaning detergents from the chamber walls and deposit

them onto instruments and wrappers.
Rust Deposits: It is very unlikely for surgical grade steel to rust. What

appears to be rust is often residual organic matter in box locks or

mineral deposits which have been baked onto the surface of the

instrument. In localities where the water has a high iron content,

for example, an iron deposit will result in a metallic film on the

instrument. This may be prevented with the use of “treated water”

for the FINAL rinse during cleaning procedures.
The most effective method of dealing with instrument problems is

to prevent them from occurring. The use of “treated water”, careful

preliminary cleaning, using neutralized pH solutions, following

manufacturer’s instructions, and visual inspection, will help to

keep instruments performing accurately and cosmetically free of

troublesome stains. It is important to act quickly should a problem

arise. Delay will compound the problem and irreparable harm may

result.

RETURNED GOODS POLICY

Products must be returned in unopened packages with

manufacturer’s seals intact to be accepted for replacement or

credit unless returned due to a complaint of product defect.

Determination of a product defect will be made by Miltex. Products

will not be accepted for replacement if they have been in the

possession of the customer for more than 120 days.

REPAIRS AND MAINTENANCE

Should your instruments require repair or maintenance, contact

Miltex for return authorization and address. Instruments returned

to Miltex for repair must have a statement testifying that each

instrument has been thoroughly cleaned and sterilized. Failure

to supply evidence of cleaning and disinfection will result in a

cleaning charge and delayed processing of your instrument repair.

PRODUCT INFORMATION DISCLOSURE

MILTEX, INTEGRA AND MANUFACTURER EXCLUDE ALL

WARRANTIES, WHETHER EXPRESSED OR IMPLIED, INCLUDING

BUT NOT LIMITED TO, ANY IMPLIED WARRANTIES OF

MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

NEITHER MILTEX, INTEGRA NOR MANUFACTURER SHALL BE

LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL LOSS,

DAMAGE, OR EXPENSE, DIRECTLY OR INDIRECTLY ARISING

FROM USE OF THIS PRODUCT. NEITHER MILTEX, INTEGRA NOR

MANUFACTURER ASSUME NOR AUTHORIZE ANY PERSON TO

ASSUME FOR THEM ANY OTHER OR ADDITIONAL LIABILITY OR

RESPONSIBILITY IN CONNECTION WITH THESE PRODUCTS.

SYMBOLS USED ON LABELING

Manufacturer

Authorized Representative in the European Community

Catalog number

Lot Number

See instructions for use

Consult instructions for use

Product complies with requirements of directive 93/42/

EEC for medical devices

LOT

LOT

REF

Integra York PA, Inc.

589 Davies Drive

York, PA 17402

toll free phone: 866-854-8300

phone: 717-840-9335

fax: 717-840-9347

PAD-IFU-GEN Rev. B 06/13

Padgett is a registered trademark of Integra LifeSciences Corporation.

The Integra wave logo is a trademark of Integra LifeSciences Corporation.

©2006 Integra LifeSciences Corporation. All Rights Reserved.

Integra LifeSciences Services (France) SAS Immeuble Sequoia 2
97 allée Alexandre Borodine
Parc Technologique de la Porte des Alpes
69800 Saint Priest–France
Telephone: 33 (0) 4 37 47 59 10
Fax: 33 (0) 4 37 47 59 29

Advertising