How to program

1. Press , display will switch to [Pro] which means access to pro-

gramming level. Another press will make the display skip to first

programming line

(option input signal).

2. Proceed with line parameterization using keys , and

.

3. Pressing will save the parameters after having them indicated

for short. The device will automatically exit the programming level.

4. Programming access can be disabled at programming level, in this

case programming lines will be read only. The device will signal

any programming lock by [dAtA] instead of [Pro] upon access.

4.1 Input configuration

Selection input signal
DC current or voltage
AC current or voltage

Voltage up to 600 V
Voltage up to 100 V
Current up to 5 A
Current up to 1 A

4.2 Display configuration

Scaling mode
SCAL mode (keypad mode)
Teach-in mode

First measured value
Manual entry in SCAL mode.

In Teach mode, the value of the provided analog signal will be

adopted.

Assigning the first measured value a displayed value
Requires manual value entry and assignment using the keypad.

Admissible range:

DC: from –1999 to 9999

AC: 0 to 9999

Decimal point
Position of decimal point (related to dSP1).