Installation of 3A approved and EHEDG certified products:

1) Use only a 3A approved counter part

2) The inspection hole should be visible and drained

3) Mount the adapter in a self drained position/angle

4) Level the inner surface of the pipe with the counter part.

After installation

Check the leak tightness of the sleeve.

Check the tightness of glands or M12 plugs.

Check the tightness of the cover

Mounting

Example
with PM060-2

Example with
PM023 or PM053

Example
with PM060-1

Leakage indication hole

must be places downwards

Fitted
incorrect

Cannot be
drained

WARNING

5) The 3A mark or the arrow shall be placed upwards

6) Welding should be grinded to Ra= 0.8

7) Tighten the swirling nipple on the AFIx 20...25 Nm

8) Installed with correct torque the electrical connection

will point downwards

≥1

m

et

er

The sensor must be fully submersed in the media.

To avoid turbulence problems the sensor is recommended to mount in
a distance of >1 meter from a bend.

The hole through the sensor must be in the flow direction

The marks on the
connection must
point in flow direction

There must be some bi-directional flow through the hole to ensure a
homogeneous media - i.e. correct measurement. There must be no
air-bubbles in the hole.