MS-(H)-S Analog (Hotplate) Magnetic Stirrer

1 Safety Instructions

Warning!

• Read the operating instructions carefully before

use the instrument.

• Ensure that only trained staff work with the

instrument.

• Forbid to heat the substances with low- burning

point or easy-volatile(MS-H-S)

Risk of burn!

• Caution when touching the housing parts and

the heating plate. The heating plate can reach

temperatures of 340 ℃.

• Pay attention to the residual heat after switching

off.

Protective ground contact!

• Make sure that socket is earthed (protective

ground contact) before use.

• When work ,wear the personal guard to avoid the risk

from:

- Splashing and evaporation of liquids

- Release of toxic or combustible gases.

• Set up the instrument in a spacious area on an stable,

clean, non-slip, dry and fireproof surface, do not operate

the instrument in explosive atmospheres, with hazardous

substances or under water.

• Gradually increase the speed, reduce the speed if :

- The stirring bar breakaway because of too high speed

- The instrument is not running smoothly, or if the

container moves on the base plate.

• Temperature must always be set to at least 25 ℃ lower

than the fire point of the media used.

• Beware of hazards due to:

- Flammable materials or media with a low boiling

temperature

- Overfilling of media

- Unsafe container

• Process pathogenic materials only in closed vessels.

• If the case of the stirrer bar is PTFE,please note:

- Elemental fluorine, three fluoride and alkali metals will

corrode the PTFE and Halogen alkanes make it expansion at

room temperature

- Molten alkali ,alkaline earth metals or their solution, as

well as the powder in second and third ethnic of the periodic

table of elements will have chemical reaction with PTFE

when temperature reaches 300 ~ 400 ℃.

• The voltage stated on the label must correspond to the

main power supply.

• Ensure that the mains power supply cable does not touch

the hotplate. Do not cover the instrument.