Faults, Maintenance and cleaning – Scilogex SK-D3309-Pro User Manual

Page 10

Advertising
background image

 

6

 

Rocking Shaker & 3D Shaker

6. Faults

When switched ON, the instrument does not work.
„ Check whether the power cable is connected.
„ Check whether the fuse is broken.

Shaking function suddenly stop.
„ LCD displays “Er 03”, indicate the current failure

is “overload protection”.

„ Pressing the speed control knob or timer control

knob first to close shaking function, and then

restart shaking function after lowered load. If

shaking function cannot start, then repeat the

process and gradually reduce load.

7. Maintenance and Cleaning

Proper maintenance can keep instrument working

properly and lengthen its lifetime.

Do not spray cleanser into the instrument when

cleaning.

Do not remove the power line when cleaning. Only

use recommended cleansers:

Dyes Isopropyl

alcohol

Construction

materials

Water containing tenside /

Isopropyl alcohol

Cosmetics

Water containing tenside /

Isopropyl alcohol

Foodstuffs

Water containing tenside

Fuels

Water containing tenside

Before using other method for cleaning or

decontamination, the user must ascertain with the

manufacturer that this method will not damage the

instrument. Wear the proper protective gloves during

cleaning of the instrument.

 

Note:

Electronic device cannot clean with cleanser.

If you require maintenance service, must be

cleaned the instrument in advance to avoid

pollution of hazardous substances, and to

send back into original packing.

If the instrument will not use for a long time,

please switch off and place in a dry, clean,

room temperature and stable location.