Operation, Faults – Scilogex MX-RL-Pro User Manual

Page 10

Advertising
background image

 

6

 

Rotator

5. Operation

Place the Rotator in safe and stable surface, ensure the

required operating voltage and power supply voltage

matched, and the socket must be earthed reliably.

Connect the power cable.

Switch ON instrument.

The instrument begins self-checking.

When initialization is over, LCD displays the last

running values of speed.

Rotate the speed/timer control knob to set target speed.

a.

Press the speed/timer control knob to start

rotating function. Press the knob again to stop rotating

function.

b.

Press the mode switch button, LCD displays time,

and then rotating the speed/timer control knob to set

target time. Press the speed/timer control knob to start

rotating function at the set time. During operation,

motion can be stopped at any time by pressing the

speed/timer control knob. If the knob is pressed again,

motion will start again and the timer will

restart

countdown. When the timer reaches zero, the unit will

be automatically halted and an alert will sound.

If these operations above are normal, the instrument is

ready to operate. If not, the instrument may be damaged

during transportation, please contact technical support of

manufacturer/supplier.

Note:

The set values can be altered at any time.

6. Faults

Instrument cannot be powered ON when start rotating

function.
„ Check whether the power cable is connected.

Rotating function suddenly stop
„ Power/overload protection light changed to red,

LCD displays “Er 03”, indicate the current failure

is “overload protection”.

„ Pressing the speed/timer control knob first to

close rotating function, and then restart rotating

function after lowered load. If rotating function

cannot start, then repeat the process and gradually

reduce load.

Advertising