Taylor-Wharton XL-240 User Manual

Page 8

Advertising
background image

Filling the Container by Volume

1.

Visually inspect the container. Do not attempt to fill containers that have broken or
missing components.

2.

Connect a transfer hose to the LIQUID fittings from a low-pressure source of liquid.

3.

Open the supply valve. Then, on the XL-160/XL-180, open the LIQUID and VENT
valves to begin the fill.

4.

When liquid begins to spit from the VENT valve, quickly close the LIQUID valve
and then the VENT valve. Both valves must be closed before the container relief
valve opens.

5.

Open the dump valve on fills line assembly to vent the fill line assembly.

6.

Disconnect the fill line from the container.

Fill Hose Kits

Taylor-Wharton fill hose kits for the XL-160/XL-180 are designed to transfer liquefied
gases to, or from the containers. These accessories are comprised of a Fill Tee Assem-
bly and a Fill Hose. Cryogenic transfer hoses are constructed of stainless steel suitable
for cryogenic liquids, and are available in 4 or 6-ft. (1.2 or 1.8 m) lengths with a 3/8 in.
NPT fitting on one end and CGA service specific female fittings on the other. A Fill Tee
Assembly consists of a cross fitting with a CGA end fitting, relief valve and manual
dump vave.

Pressure Transfer Filling

Fill Hose Kits