1 sample program for cr200(x) series datalogger, 2 example 2. sample program for cr1000 datalogger – Campbell Manufacturing TEMPERATURE PROBE 109SS User Manual

Page 10

Advertising
background image

Model 109SS Temperature Probe

5.1.1.1 Sample Program for CR200(X) Series Datalogger

'CR200(X) Series Datalogger

‘This example program measures a single 109SS Thermistor Probe
‘once a second and stores the average temperature every 10 minutes.

‘Declare the variable for the temperature measurement
Public Air_Temp

‘Define a data table for 10 minute averages:
DataTable (AvgTemp,1,1000)
DataInterval

(0,10,min)

Average

(1,Air_Temp,0)

EndTable

BeginProg

Scan (1 ,sec)

‘Measure the temperature:

Therm109

(Air_Temp,1,1,Ex1,1.0,0)

Call the data table:

CallTable

AvgTemp

NextScan
EndProg

5.1.1.2 Example 2. Sample Program for CR1000 Datalogger

'CR1000

'Declare Variables and Units
Public T109_C

Units T109_C=Deg C

'Define Data Tables
DataTable(Table1,True,-1)
DataInterval(0,10,Min,10)
Average(1,T109_C,FP2,False)
EndTable

'Main Program
BeginProg
Scan(1,Sec,1,0)

'Default Datalogger Battery Voltage measurement Batt_Volt:

'109SS Temperature Probe measurement T109_C:

Therm109(T109_C,1,1,1,0,_60Hz,1.0,0.0)

'Call Data Tables and Store Data

CallTable(Table1)
NextScan
EndProg

6