Salter Brecknell C3235 User Manual

Page 12

Advertising
background image

12

2.

Place the scale on stable, level surface and turn the unit on.

3.

After the self-check is complete, press and hold the TARE key
until…

CAL is briefly displayed, then 0.

4.

Empty the scale and press the TARE key to set the zero point.

Hg or Lb is displayed.

5.

Toggle between the choice of calibration weight unit of measure
using the F key. When the unit of measure you want to use is
displayed, press the TARE key…

The full capacity will flash.

6.

Load the full capacity weight onto the scale platform and press the
TARE key.

CAEnd (calibration end) will briefly display, then the unit returns
to normal weigh mode.

If you replace the PCB or loadcell, the scale requires recalibration. If the
display shows Err-0 after power up, it means there is a problem with
calibration. Press and hold the TARE key to enter calibration and repeat
the calibration process.

Multipoint Calibration:

Follow these steps for multipoint calibration of the scale. You will need test
weights equal to 1/3, 2/3 and full capacity.

1.

Place the calibration switch, located on the bottom of the scale into
the ON position. You may need to remove the small cover to access
the calibration switch. Use a small screwdriver to remove the two
screws holding the cover.

2.

Place the scale on stable, level surface and turn the unit on.

3.

After the self-check is complete, press and hold the F and ZERO
keys until…

20 or 100 is displayed. This stands for the powerup zero range.

4.

Repeatedly press the F key to toggle between the two choices: 20
and 100. When your choice is displayed, press the TARE key…

Lb or Kg is displayed. This is the calibration unit of measure.

5.

Press the F key to toggle between the two choices. When your
choice is displayed, press the TARE key…

Scale full capacity is displayed.